Standard applications in process and environment Long-term monitoring of quiet stable processes
Characteristic
Digital, contactless electrode with Memosens technology Very robust, heavy-duty applications, high pressure version
Measuring range
-1500 mV to +1500 mV
Measuring principle
Gel compact electrode with teflon diaphragm and double gel reference Measuring part as gold-pin or platin ring
Design
All shaft lengths with temperature sensor Advanced gel technology Inductive, digital data transfer with Memosens technology
Material
Glass electrode with teflon diaphragm
Dimension
Diameter: 12 mm (0.47 inch) Shaft lengths: 120, 225 and 360 mm (4.72, 8.86 and 14.17 inch)
Process temperature
-15 to 135 °C (5 to 275 °F)
Process pressure
0 to 16 bar (0 to 232 psi)
Temperature sensor
NTC 30K
Ex certification
ATEX II 1G EEX ia IIC T3/T4/T6 FM/CSA Class 1 Div.2, in combination with CM42 and CPM153 transmitters
Connection
Inductive, digital connection head with Memosens technology
Ingress protection
IP68
Additional certifications
Biocompatibility validated according to ISO 10993-5:1993 and USP(current version) TÜV certificate for pressure resistance up to 16 bar with triple overpressure safety Electromagnetic compatibility complies with EN 61326:1997
Technical Information Orbisint CPS12/ CPS12D / CPS13
Technical Information (TI)
English version - 05/2005
New version available in English
Redox electrodes, analog and digital with Memosens technology and reference electrode Each with dirt-repellent PTFE diaphragm forstandard applications in process and environmental technology
Memosens pH/ORP
Sicherheitshinweise für elektrische Betriebsmittel im
explosionsgefährdeten Bereich
Safety instructions for electrical apparatus for explosion-hazardous
areas
Conseils de sécurité pour matériels électriques destinés aux zones
explosibles
Safety Instructions
Ex Safety Instruction (XA)
English version - 11/2016
New version available in English
Safety instructions for electrical apparatus for explosion-hazardous areas
Memosens pH/ORP
Sicherheitshinweise für elektrische Betriebsmittel im
explosionsgefährdeten Bereich
Safety instructions for electrical apparatus for explosion-hazardous
areas
Conseils de sécurité pour matériels électriques destinés aux zones
explosibles
Safety Instructions
Ex Safety Instruction (XA)
English version - 02/2016
New version available in English
Safety instructions for electrical apparatus for explosion-hazardous areas
pH and ORP measurement Supplement to: BA01572C, BA01109C Safety instructions for electrical equipment in explosion-hazardous areas EAC Ex 0Ex ia IIC T3/T4/T6 Ga X EAC Ex 0Ex ia IIC T4/T6 Ga X
pH/ORP measurement Supplement to BA01572C, BA01109C Safety instructions for electrical apparatus in explosion-hazardous areas UKEx II 1 G Ex ia IIC T3/T4/T6 Ga UK Ex II 1 G Ex ia IIC T4/T6 Ga